Configure Gas Inlets
For each attached calibration gas cylinder, you must configure the gas blend, and then configure each gas type and concentration ontained in the gas blend. You may also record the lot number of the calibration gas cylinder
If a single-inlet key is inserted in an individual module or all connected modules in a gang, then you can attach one calibration gas cylinder. If a multi-inlet key is inserted in an individual module or one or more connected modules in a gang, then, then you can attach up to four calibration gas cylinders. For more information, see Inlet Keys on page 9
If the module is part of a gang of up to 5 connected modules, then changes to the gas inlet settings affect all of the modules in the gang. If the gang contains both single-inlet and multi-inlet key modules, then changes to Inlet 1 affect all modules and changes to Inlets 2, 3 and 4 affect the multi-inlet key modules only.
3. The Inlet overview for Inlet 1 is displayed. If a multi-inlet key is inserted in an individual module or one or more connected modules in a gang, then four inlet tabs are displayed. Press orto select and configure a different inlet.
4. Configure the gas blend for the selected inlet.