1-866-4 JJSTECH(1-866-455-7832)

E Instruments E4400-3D Industrial Combustion Gas & Emissions Analyzer with O2, CO, NO/NOx Gas Sensors and CO Dilution Auto-Range to 50,000 ppm

Item #:
E4400-3D
Brand:
E Instruments
E Instruments E4400-3D Industrial Combustion Gas & Emissions Analyzer with O2, CO, NO/NOx Gas Sensors and CO Dilution Auto-Range to 50,000 ppm
$2,995.00
Qty

Discontinued

Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
E Instruments E4400-3D Industrial Combustion Gas & Emissions Analyzer with O2, CO, NO/NOx Gas Sensors and CO Dilution Auto-Range to 50,000 ppm

E Instruments E4400-3D Industrial Combustion Gas & Emissions Analyzer with O2, CO, NO/NOx Gas Sensors and CO Dilution Auto-Range to 50,000...

$2,995.00

Discontinued


E Instruments E4400 Industrial Combustion Gas and Emissions Analyzer

The E4400 is a new handheld combustible gas and emissions analyzer by E Instruments International. The device incorporates the best elements of previous models while incorporating modern and advanced technology. With an emphasis on customer needs, the E4400 is an incredible tool for onsite safety management.

Exclusive Features

The E4400 Industrial Combustion Gas and Emissions Analyzer uses up to four different gas sensors with low NOx and true NOx capabilities. The device has a state of the art non-thermal printer for onsite gas analyzing and assessing and a draft and differential pressure meter. With its 2 channel thermometer, instant response time, and rechargeable lithium ion battery, the analyzer is reliable for consistent field use. To improve its durability and life span, the unit also features an external condensation trap and unbreakable metal probe and hose connections. The unit has a magnet in it allowing for complete hands free operation. Each device has software included that provides unlimited storage for data.

E Instruments E4400 Industrial Combustion Gas and Emissions Analyzer Specifications

The unit device is compatible with a variety of sensors to monitor CO, CO2, O2, NO, NO2, NOx, SO2, CxHy, efficiency and excess air, Tair Tgas, and draft and pressure. The sensors are pre-calibrated and easy to install when in the field. Its unbreakable metal hose connectors are also conducive to harsh field work. It is useful for working with a variety of industrial combustible equipment such as

  • Boilers
  • Heaters
  • Burners
  • Kilns
  • Engines
  • Furnaces
  • Turbines

Product Inclusions

Each E4400 model includes: a built-in impact printer; draft/pressure and temperature measurements; calculations for CO2, efficiency and excess air; gas sampling probe and hoses, rechargeable batteries and AC charger, water trap with filter cartridge, internal memory, PC software package, durable rubber casing with magnetized supports, USB cable, calibration certificate, operator’s manual, and an ABS hard plastic carrying case.

E Instruments International

E Instruments International is a developer, manufacturer, and distributor of innovative instrumentation solutions. Their product offerings include a variety of air quality analyzers and calibration equipment. E Instruments International products are designed specifically for certain industries such as the power, process, industrial, institutional, food, and HVAC industries. Over the years their research and development team has developed a reputation for creating and delivering quality and cost effective safety solutions to meet consumer demands. The company is also dedicated to customer support and applications. The customer support team is always available to meet consumer needs and answer questions about products.

Parameter
Sensor
Range
Resolution
Accuracy
O2 Electrochemical 0 - 25% 0.1% ±0.2% vol
CO H2 Compensated w/ Built-in NOx Filter Electrochemical 0 - 8000 ppm 1 ppm ±10 ppm (0-200ppm)
±5% rdg (201-2000ppm)
±10% rdg (2001-8000ppm)
CO DILUTED Electrochemical 0.8% - 5.00% 0.01% ±10% rdg
CO2 Calculated 0 - 99.9% 0.1%  
NO Electrochemical 0 - 5000 ppm 1 ppm ±5 ppm (0-100ppm)
±5% rdg (101-5000ppm)
NO2 Electrochemical 0 - 1000 ppm 1 ppm ±5 ppm (0-100ppm)
±5% rdg (101-1000ppm)
Low NO and/or Low NO2 Electrochemical 0 - 500 ppm 0.1 ppm ±2 ppm (0.0-40.0ppm)
±5% rdg (40.1-500.0ppm)
NOx Calculated 0 - 5000 ppm 1 ppm  
SO2 Electrochemical 0 - 5000 ppm 1 ppm ±5 ppm (0-100ppm)
±5% rdg (101-5000ppm)
CxHy Pellistor 0 - 5 % 0.01 % ±5 % full scale
Tair Pt100 -10 to 212°F
-10 to 100°C
0.1°F
0.1°C
± 1°F
± 1°C
Tgas Tc K -4 to 2280°F
-20 to 1250°C
0.1°F
0.1°C
±0.5°C (-20 to 100°C)
±0.5% rdg (101 to 1250°C)
Differential Temp. Calculated -10 to 2280°F 0.1°F  
Pressure/Draft Semiconductor + 0 - 80 inH2O ±1.0% rdg  
Excess Air Calculated 0 - 850% 1 %  
Efficiency Calculated 0 - 100% 0.1%  

Datasheets & User Manuals

Questions & Answers


There are no questions or answers for this item.